Custom Dental Suction Systems

At Becker Pumps Australia we recognise three main types of Dental Suction Systems that are used by Dental Surgeries, Dental Hospitals and Training Facilities;

  1. Dry (line) Suction System; this system removes solids and liquids within the surgery leaving the pipeline from the separator to the suction pump dry.
  2. Semi-Dry (line) Suction System; this system removes solids from one or a number of surgeries at a central collection point adjacent to the surgery or surgeries.  The pipeline from the separator to the suction pump is dry.
  3. Wet (line) Suction System; this system discharges to waste solids less than 600mm mesh size and liquids through a "wet" pipeline via a suction pump.

At Becker Pumps Australia we offer both the Dry and Semi-Dry Dental Suction Systems which meet the requirements under the discontinued Australian Standard AS 2686.1-1984 as well many of the current standard standards from around the world.

Becker Pumps Australia can supply one of our standard units or custom design and build a unit to suit your requirements.  We have supplied systems ranging from one chair up to 150 chairs; but we are not limited at that number. 

Currently we have six (6) standard Dental Suction Systems (3 x Dry and 3 x Semi-Dry) and three (3) Separator and Waste Pump Systems to suit our Dry Systems.

  • The Semi-Dry Systems consist of a rack mounted system complete with a Becker vacuum pump, control box, bacterial filter, separator and waste pump.
  • The Dry Systems consist of a rack mounted system complete witha Becker vacuum pump, control box and bacterial filter.
  • The Separator and Waste Pump Systems consist of a skid mounted separator and waste pump and are sized to match our Dry Systems.

All of our Dental Suction System are fitted with a Becker Side Channel Vacuum Pump.  Whilst the pumps that we use have maximum pressure greater than required in dental suction, our pumps are fitted with relief valves to vent and maintain the systems automatically at a gauge pressure of not more than -20 kPa.

To see the specifications of our standard systems, click here.

The Becker Pumps Australia Dental Suction Systems

MEETS STANDARDS Although the Australian Standard AS 2686.1-1984 is no longer current, our Dental Suction Systems meet with all the requirements of that standard. 

150 Chair Suction System

AUSTRALIAN ASSEMBLY Our systems are manufactured using German manufactured Becker pumps, and your system is assembled in Australia to suit your requirements.
ENERGY SAVINGS Using our Variable Frequency Driven (VFD) Pumps, the energy usage is matched to the demand of the system therefore saving running costs when low demand is required.
CASE STUDIES Please visit our Case Studies page for a list of Dental Suction Systems supplied by Becker Pumps Australia and installed throughout the country.

The Vacuum Pumps

SIDE CHANNEL VACUUM PUMPS We install our 100% oil-free vacuum pumps which are regarded as the world's most popular oil-free pumps in many applications.

Becker Variair Side Channel Pump

 

HIGH EFFICIENCY MOTORS

Our pumps use heavy duty high efficiency motors with class-F insulation as standard equipment. The pumps are direct driven with a single shaft meaning there are no gear boxes or twin shafts which are known to have common failures.
PATENTED VFD TECHNOLOGY On our systems with the variable frequency drive technology, we use the Becker Variair model pumps.  Variair is Becker's Patented Intelligent Variable Speed Drive.  To find out more about our Variair pumps, click here.
NO VIBRATION Smooth running side channel pumps require no special foundations; Vibration meets AS 2625.
QUIET Rotary operation has no pulsations, and sound reducing technology is built-in, resulting in some of the quietest pumps in the world. The noise levels meet AS/NZS 2107 for health buildings.
AIR COOLED All Becker pumps are air cooled, meaning no water is ever needed for cooling and operation.
NO WATER None of our pumps use water for any reason. Water and sewer charges are eliminated and water pollution is never a problem. Furthermore, your dependence on another utility is eliminated.
100% OIL FREE There is no oil to fill, drain, change, or discard. No oil aerosols are ever present in the discharge air.
INLET FILTERS Every system includes inlet filters before each pump to prevent pump damage due to foreign particles.

The Benefits of the Becker Pumps Australia Dental Suction Systems

ENERGY SAVINGS Using our Variable Frequency Driven (VFD) Pumps, the energy usage is matched to the demand of the system therefore saving running costs when low demand is required.

DSV-DRY 2-6

DS-SWP 40L

SUCTION Constant suction is always available.
FAIL SAFE Our complete control failure safety system in multiple pump systems has fail safe capabilities.  In the even of controller failure, all pumps will revert to 60Hz.
LOW MAINTENANCE Longer service intervals, reduced wear, no vanes to change, no oil disposal, and easy to clean.  One central point for separation and cleaning.
CONTROL SYSTEM Simple, easy to use controller with full connectivity with alarm panels. BACnet gateway available.
COMPACT & MODULAR Whilst our systems have a small footprint, they are modular allowing for the system to expand when the facility (hospital, surgery or training facility) grows.
FILTRATION Our system's filtration complies with ISO 8573-1 Class 1.
AGSS GAS SCAVENGE Scavenged oxygen safe.
INSTALLATION SAVINGS There is no need to elevate the plant room therefore building costs are reduced.
SMALLER RECEIVER When using our VFD pumps on the larger custom designed and built systems, a smaller receiver can be used as per AS 2896-2011.

The Inclusions

THE PACKAGE Approved industrial quality electronic control, built in interface connection with air/water separator (complete with our Semi-Dry systems), remote stop/start, hour meter, and vacuum gauge.

DSV-SEMI 2-6

AIR/WATER SEPARATORS These are included on all Standard and Custom Designed and Built Semi-Dry Dental Suction Systems with three (3) options available to be used in conjunction with our Dry Dental Suction Systems (refer below).
MANUALS All Becker Pumps Australia Dental Suction Systems come complete with full manuals supplied in a hardcopy and soft copy.  The manual includes the operating and maintenance instructions for the pumps, details about the PLC installed along with the key features it has been programmed to do, and other useful information about all the equipment as installed in your system.

 

The Becker Pumps Australia Standard Dental Suction Systems

 

SYSTEM

TYPE

NUMBER

OF CHAIRS

POWER*

(kW)

PUMP

INSTALLED

NOISE LEVEL*

(dbA)

DIMENSIONS

(mm)

SEPARATOR

SIZE (litres)

DSV-SEMI 2-6 60L

Variable Speed

Semi-Dry

2 to 6 4.0 SV 8.130/1-401 74

length 1150

width 600

height 1100

60

DSV-SEMI 7-10 60L

Variable Speed

Semi-Dry

7 to 10 4.0 SV 8.190/1-401 69

length 1150

width 600

height 1100

60

DSV-SEMI 11-20 120L

Variable Speed

Semi-Dry

11 to 20 7.5 SV 7.330/1-401 76

length 1150

width 800

height 1500

120

DSV-DRY 2-6

Variable Speed

Dry

2 to 6 4.0 SV 8.130/1-401 74

length 600

width 600

height 1150

not included

refer table

below

DSV-DRY 7-10

Variable Speed

Dry

7 to 10 4.0 SV 8.190/1-401 69

length 600

width 600

height 1150

not included

refer table

below

DSV-DRY 11-20

Variable Speed

Dry

11 to 20 7.5 SV 7.330/1-401 76

length 600

width 800

height 1550

not include

refer table

below

CUSTOM

Variable Speed

Semi-Dry or Dry

20 plus

Multiple pumps chosen

to match the design flow

based on chair numbers

Between

69 to 74

Sized to meet

allocated space

in plant room

sized to meet

the demands

of chair numbers

NOTES

All of the above systems require 400-480V Three Phase Power.

All pumps have maximum pressure greater than -20kPa with relief valves limiting them to -20kPa.

 

Air/Water Separators and Waste Pump Systems

The Becker Pumps Australia Semi-Dry Dental Suction Systems are supplied with Air/Water Separators and Waste Pump as standard . 

The Dry Dental Suction Systems require a separate Air/Water Separator and Waste Pump System.  These are matched perfectly to the recommended system shown below, but each system can be mixed and matched and multiple smaller Air/Water Separator and Waste Pump Systems can be used on the larger dental Suction Systems.


SEPARATOR

TYPE

CAPACITY

(litres)

INLET

(inches)

OUTLET

(inches)

WASTE PUMP

MOTOR (kW)

DRY SYSTEM

MATCHED TO

DIMENSIONS

(mm)

DS-SWP 40L

Air/Water Separator

c/w Waste Pump

40

2" BSP

1" 0.37

DSV-DRY 2-6

Length 980

Width 500

Height

DS-SWP 60L

Air/Water Separator

c/w Waste Pump

60 2" BSP 1" 0.37

DSV-DRY 2-6

DSV DRY 7-10

Length 980

Width 500

Height

DS-SWP 120L

Air/Water Separator

c/w Waste Pump

120 2½" BSP 1" 0.37 DSV-DRY 11-20

Length 980

Width 500

Height

NOTES All of the above systems require 240V single phase power.

 

 

 

Home
About Us
Contact Us

Facebook
Twitter
You Tube

Pumps

Document – Downloads
Data Sheets – Downloads

Vacuum Pumps

Oil-Free Rotary Vane
Oil-Lubricated Rotary Vane
Screw Pumps
Side Channel
Radial

Compressors

Oil-Free Rotary Vane
Oil-Lubricated Rotary Vane
Screw Pumps
Side Channel
Radial

Combined Pumps

Oil-Free Rotary Vane
Side Channel

Central Systems

Service

Operating Manual – Downloads
Hire Pumps
Service Agreement Download

Medical

Medical suction system downloads
Medical Suction Systems
Dental Suction System
Anesthetic Gas Scavenge Suction Systems
Laboratory Suction System

Spare Parts

Spare Parts Lists – Downloads

Accessories

External Filters and Separators
Lubricants – Oil and Grease
Pressure Gauges
Sound Proof Boxes
Receivers (Air Tanks)
Change-Over Valves
Non-Return Valves

 

You Tube Twitter Facebook Bookmark Site Tell A Friend Print Contact Us Home1300 579 177 Request A Quotation